Information
Specifications
Information
Small footprint, saving precious laboratory space.
Unique design utilizing a dual-valve gradient for cost effective gradient solutions.
Integrated multifunction valves offer flexibility and automatic path switching.
Enhanced process efficiency when used with the next generation PURfect P® Protein Purification System Software.
Specifications
Parameter | SEP 25 | |||
---|---|---|---|---|
Pump | ||||
Type | Binary Piston Pump | |||
Flow Rate | 0.01-25ml/min | |||
Pressure | 0-10MPa(100bar,1450psi) | Flow Rate Accuracy | ±1.2% | |
Valve | ||||
Inlet | 1 Sample inlet and 3 mobile phase inlets ( Two mobile phase inlets used for gradient elution ) | |||
Automated Injection | Supports Manual Load, Injection, and Waste | |||
Column Position ( Optional ) | Single valve with 3 column positions; Column, Bypass, and reverse flush functions | |||
Collection ( optional ) | 8 outlet options F1-F8 | |||
UV Detector | ||||
Wavelength Options |
|
|||
Wavelength Precision and Repeatability | ±1nm / ±0.5nm | |||
Noise/Shift | 0.16mAu (1s) / 1mAu/h | |||
Optical Path | 2mm | |||
Conductivity Detector | ||||
Detection Range | 0.001-999.99mS/cm | |||
Detection Accuracy | ±0.1mS/cm or ±2% | |||
Temperature Detector | ||||
Detection Range | 0-100℃ | |||
Detection Accuracy | ±1℃ | |||
pH Detector | ||||
Detection Range | 0-14 | |||
Detection Accuracy | ±0.1(2-12) | |||
System Details | ||||
Chromatography Workstation | PURfect P® brand new chromatography workstation on PC w/ Win10 system | |||
Weight | 50kg (110lbs) | |||
Dimensions | 510mm X 600mm X 580mm (length X width X height) | |||
Fraction Collector (optional) | ||||
Rack | 5ml*90, 15ml*60, 50ml*21, and 96 well plate |